C1 INHIBITOR - plasma and serum
Specimen: 5 mL
blood in plain tube (immunochemical assay) and 5 mL blood in EDTA tube (functional
assay). The specimens should be placed on melting ice and transported immediately
to the laboratory.
Method: Functional and immunochemical assay.
Reference Interval: Immunochemical 0.25-0.42 g/L; functional: consult pathologist.
Application: Levels of C3 and C4 should be determined before this test is requested;
see
COMPLEMENT COMPONENTS C3 AND C4
. Investigation of angioedema - C1 inhibitor levels are unlikely to be reduced if
C4 is normal. Diagnosis of hereditary angioedema due to C1 inhibitor deficiency,
acquired C1 inhibitor deficiency in malignancy and SLE. Genetic and acquired forms
of C1 inhibitor deficiency are distinguished by family history and on clinical grounds.
Interpretation: In genetic deficiency, 85% of kindreds have low functional and
immunochemical levels; 15% of kindreds have low functional levels with normal or
high immunochemical levels. Levels may rise with danazol treatment.
Reference: Alsenz J et al. N Engl J Med 1987; 316: 1360-1366. Davis AE.
Annu Rev Immunol 1988; 6: 595-628.